Regulatory status
This page states FerroFED’s intended purpose and how it is classified under Regulation (EU) 2025/327 on the European Health Data Space (the EHDS Regulation). The classification was decided on 2026-10-04 (#519): FerroFED is built and documented as an EHR system, and its harmonised software components are to be delivered before the dates the Regulation applies to it. The page also states FerroFED’s position under the Cyber Resilience Act, decided on 2026-10-06 (#654): each release is a product that Cadasto B.V. places on the market (The Cyber Resilience Act).
Every quotation from the EHDS Regulation is from the Official Journal text
(OJ L, 2025/327, 5.3.2025,
http://data.europa.eu/eli/reg/2025/327/oj), vendored verbatim at
docs/specs/eu-ehds/reg-eu-2025-327-en.xhtml.
This page is not legal advice; the last section names
what a deployment’s counsel should confirm.
Intended purpose
FerroFED is a federation gateway intended by its manufacturer to be used by healthcare providers when providing patient care. A clinician’s application sends it an ordinary AQL query about a patient. FerroFED resolves the patient to a record at each openEHR clinical data repository (CDR) of a federation, asks each CDR for its part of the answer, and returns one answer that names the CDR each row came from. It routes the follow-up reads and writes of that answer to the CDR that holds the record.
- Users. Healthcare providers and their health professionals, through the client applications they use in patient care. Every caller is authenticated at the gateway (client authentication). The operator of a deployment configures the member CDRs and the identity services. FerroFED offers patients no access service of their own, the “electronic health data access service” of Art 2(2)(h).
- Data. Personal electronic health data, “data concerning health and genetic data, processed in an electronic form” (Art 2(2)(a)), held in the member CDRs as openEHR records. FerroFED holds no clinical data of its own: it passes each CDR’s answer through, merges the answers, and forwards a write to the CDR that owns the record. What it keeps is the registry of members, routing state in memory, the stored-query definitions and the IHE audit records waiting for delivery; Data protection lists each.
- Priority categories. Art 14(1) lists six: “(a) patient summaries; (b) electronic prescriptions; (c) electronic dispensations; (d) medical imaging studies and related imaging reports; (e) medical test results, including laboratory and other diagnostic results and related reports; and (f) discharge reports”, with their main characteristics in Annex I. FerroFED does not select data by category: a query reaches whatever the member CDRs hold. Its intended purpose therefore covers every priority category a member CDR holds, all six of Art 14(1)(a) to (f).
Classification: an EHR system
Art 2(2)(k) defines an EHR system as:
any system whereby the software, or a combination of the hardware and the software of that system, allows personal electronic health data that belong to the priority categories of personal electronic health data established under this Regulation to be stored, intermediated, exported, imported, converted, edited or viewed, and intended by the manufacturer to be used by healthcare providers when providing patient care or by patients when accessing their electronic health data
FerroFED meets each part of that definition:
- It is software that allows priority-category data to be intermediated. Answering a clinician’s query from many CDRs, and routing a read or a write to the CDR that owns the record, is intermediation. The Regulation expects an EHR system that intermediates without storing: Annex II, point 2.1, begins “Where an EHR system is designed to store or intermediate personal electronic health data”.
- The data belong to the priority categories. Data of each of the six categories can be held in an openEHR CDR, and the intended purpose covers every category the member CDRs hold.
- The manufacturer intends it for healthcare providers providing patient care. That is the intended purpose above. Defining the purpose to exclude the priority categories would cut FerroFED off from its own use case, so the purpose includes them.
The general-purpose exclusion
Art 25(2) reads: “This Chapter shall not apply to general purpose software used in a healthcare environment.” Recital 38 gives the examples it has in mind: “text-processing software used for writing reports that would then become part of written electronic health records, general-purpose middleware, or database management software that is used as part of data storage solutions.”
FerroFED is middleware, so the recital’s second example needs an answer. It is built for one clinical purpose. It speaks openEHR ITS-REST and AQL to openEHR CDRs, it resolves the patient through an identity cross-reference, it rewrites the patient’s query for each CDR, and its intended purpose names healthcare providers providing patient care. The same recital contrasts general-purpose software with “EHR systems specifically intended by the manufacturer to be used for processing one or more specific categories of electronic health data”, and FerroFED is the second kind.
The obligations and their dates
Art 25(1) sets the core obligation: “EHR systems shall include a European interoperability software component for EHR systems and a European logging software component for EHR systems (the ‘harmonised software components of EHR systems’)”. The interoperability component “provides and receives personal electronic health data under a priority category for primary use … in the European electronic health record exchange format” (Art 2(2)(n)). The logging component “provides logging information related to access by health professionals or other individuals to priority categories of personal electronic health data … in the format defined in point 3.2. of Annex II” (Art 2(2)(o)).
Art 105 sets the dates. The Regulation “shall apply from 26 March 2027”, and Articles 25, 26, 27, 47, 48 and 49 apply later: “from 26 March 2029 to … EHR systems intended by the manufacturer to process” patient summaries, prescriptions or dispensations, and “from 26 March 2031” to those intended to process imaging, test results or discharge reports. FerroFED’s intended purpose covers all six categories, so both dates apply to it: 26 March 2029 for patient summaries, prescriptions and dispensations (Art 14(1)(a) to (c)), and 26 March 2031 for imaging, test results and discharge reports (Art 14(1)(d) to (f)).
| Obligation | Article | Applies to FerroFED from |
|---|---|---|
| Include the two harmonised software components | Art 25(1) | 26 March 2029 for (a) to (c); 26 March 2031 for (d) to (f) |
| Be placed on the market or put into service “only if they comply with the provisions laid down in this Chapter” | Art 26(1) | 26 March 2029 for (a) to (c); 26 March 2031 for (d) to (f) |
| Registration in the EU database before placing on the market or putting into service, with the results of the testing environment | Art 49(2), Art 30(1)(h) | 26 March 2029 for (a) to (c); 26 March 2031 for (d) to (f) |
| Conformity of the harmonised components with the essential requirements of Annex II and the common specifications | Art 30(1)(a), Art 36 | 26 March 2027 (see below) |
| Technical documentation with at least the elements of Annex III and the results of the European digital testing environment | Art 37, Art 40(3), Art 30(1)(c) | 26 March 2027 (see below) |
| An information sheet naming the intended purpose and the data categories, and instructions for use | Art 38, Art 30(1)(d) | 26 March 2027 (see below) |
| The EU declaration of conformity, with the content of Annex IV | Art 39, Art 30(1)(e) | 26 March 2027 (see below) |
| The CE marking of conformity, affixed “before placing the EHR system on the market” | Art 41, Art 30(1)(f) | 26 March 2027 (see below) |
| Corrective action, complaint channels, and the registers of complaints and non-conforming systems | Art 30(1)(i) to (o) | 26 March 2027 (see below) |
| Keeping the technical documentation and the declaration “for 10 years” | Art 30(3) | 26 March 2027 (see below) |
| Reporting a serious incident “not later than three days” after becoming aware of it | Art 44(7) | 26 March 2027 (see below) |
| Identifying the economic operators supplied, on request, “for 10 years” | Art 35 | 26 March 2027 (see below) |
| No claim that ascribes “functions and properties to the EHR system which it does not have” | Art 28 | 26 March 2027 (see below) |
Art 105 names no later date for Articles 28, 30, 35 to 41 and 44, so the general date of 26 March 2027 applies to them. Each of them attaches to an EHR system placed on the market or put into service, and Art 26(1), which makes compliance the condition for either, applies to FerroFED from 26 March 2029 for categories (a) to (c) and from 26 March 2031 for (d) to (f). That reading is one counsel should confirm.
The technical content comes from implementing acts the Commission has not
yet adopted. The exchange format (Art 15(1)) and the common specifications
for the essential requirements (Art 36(1)) are both due “By 26 March 2027”,
and Art 105 applies them from the same 2029 and 2031 dates by category.
A weekly job, scripts/checks/ehds-acts.sh, reads EUR-Lex and the
Commission’s “Have your say” register and files an issue for each act
adopted; the re-checks against the exchange format and the common
specifications wait in FerroHEALTH/FerroBRIDGE#1327 and #799.
Two implementing acts are adopted. Both bind the national contact points and the cross-border authentication chain, and neither sets the content of the harmonised components:
| Act | Adopted under | Applies from |
|---|---|---|
| Commission Implementing Regulation (EU) 2026/2083 of 18 September 2026 on MyHealth@EU (OJ L, 21.9.2026) | Art 23(4) and (8) | 26 March 2027 |
| Commission Implementing Regulation (EU) 2026/2099 of 21 September 2026 on the cross-border identification and authentication mechanism (OJ L, 22.9.2026) | Art 16(2) | 26 March 2027, and its Art 3(3) and 5(2) from 26 March 2029 |
Implementing Regulation 2026/2083 has the national contact points exchange data “in accordance with the requirements catalogue” and “the technical specifications” the MyHealth@EU steering group approves (its Art 4(1)). Implementing Regulation 2026/2099 has the entity a Member State lists authenticate a health professional at eIDAS assurance level “substantial”, and at level “high” from 26 March 2032 (its Art 6(3)). Both are vendored beside the Regulation, with Commission Recommendation (EU) 2019/243 on a European Electronic Health Record exchange format, which recital 26 names as the format’s foundation. A weekly check reads EUR-Lex and the Commission’s “Have your say” register for the acts still pending, and files an issue when one is adopted.
A deployment a health institution runs for itself
Art 26(2) reads: “EHR systems that are manufactured and used within health institutions established in the Union, as well as EHR systems offered as a service … to a natural or legal person established in the Union, shall be considered as having been put into service.” Putting into service is “the first use, for its intended purpose, in the Union of an EHR system covered by this Regulation” (Art 2(2)(l)). Recital 44 explains the intent: a healthcare provider “developing and using an EHR system ‘in-house’” should “comply with all requirements applicable to manufacturers”, after “an extended transitional period”. Art 105 sets that period: “Chapter III shall apply to EHR systems put into service in the Union referred to in Article 26(2) from 26 March 2031.”
For FerroFED this means:
- A health institution that builds and runs its own FerroFED may be the one that manufactured and used it within the institution. It would then carry the manufacturer’s obligations for that system, from 26 March 2031.
- A deployment that changes FerroFED can make its operator a manufacturer whatever the answer to the first point. Art 34 treats a user as a manufacturer where it modifies an EHR system “in such a way that conformity with the applicable requirements might be affected”, or “in such a way that it leads to changes in the intended purpose declared by the manufacturer”. FerroFED’s source is open to modification (licensing), so this applies to any deployment that runs changed code.
- An operator that hosts FerroFED for other organisations offers an EHR system as a service, which Art 26(2) also counts as putting into service. Such a hosted service also needs a commercial licence (licensing).
FerroFED plans its harmonised components for before the dates that apply to each category (26 March 2029 for (a) to (c)) whichever reading holds, so a deployment can use them under either. Who carries the manufacturer’s obligations for a given deployment is a question for that deployment’s counsel.
Cross-border access
Art 11(2): where the Member State of affiliation and the Member State of treatment differ, “cross-border access to the personal electronic health data of the natural person under treatment shall be provided through the cross-border infrastructure referred to in Article 23”. That infrastructure is MyHealth@EU and each Member State’s national contact point for digital health, “an organisational and technical gateway for the provision of services linked to the cross-border exchange of personal electronic health data” (Art 23(2)). The exchange between contact points “shall be based on the European electronic health record exchange format” (Art 23(3)), and “Member States shall ensure the connection of all healthcare providers to their national contact points for digital health” (Art 23(5)). Art 105 applies Art 23(2) to (6) from the same 2029 and 2031 dates.
FerroFED does not reach another Member State itself and does not build the contact point’s cross-border protocols. Its route to cross-border care is the national one: a national contact point asks its national infrastructure for a patient’s documents, and FerroFED can answer that request from federated openEHR data through its interoperability component. It does so on its FHIR face, where a national connector reads the patient summary, the document list and the prescriptions and posts a dispensation (A national contact point, #524).
The Federation Tier specification does not say whether a federation whose member CDRs sit in different Member States is in scope. FerroFED neither detects nor refuses one; whether such a deployment is lawful, and for which data, is the deployment’s to answer.
The Cyber Resilience Act
Regulation (EU) 2024/2847, the Cyber Resilience Act (CRA), “applies to
products with digital elements made available on the market” whose intended
purpose includes a data connection to a device or network (Art 2(1)).
FerroFED is software whose purpose is that connection. The quotations in this
section are from the Official Journal text (OJ L, 2024/2847, 20.11.2024),
vendored with Regulation (EU) 2019/1020 at
docs/specs/eu-cra/,
and Art 104 of the EHDS Regulation amends three of its articles.
The manufacturer’s side of the CRA is the same for every FerroHEALTH product and is in the family book: who the manufacturer is, which duties apply from when, the vulnerability-handling policy, the Annex II information every product shares, and the conformity route of each product (The Cyber Resilience Act, the manufacturer). This section is FerroFED’s own reading: why a FerroFED release is a CRA product, what that asks of it, and its support period.
Why the CRA reaches a release
- A commercial activity. Art 3(22) defines making available on the market as “the supply of a product with digital elements for distribution or use on the Union market in the course of a commercial activity, whether in return for payment or free of charge”. Recital 15 names “an intention to monetise” as one mark of a commercial activity. FerroFED’s licence is free for non-production use and for non-commercial production use, and sells a licence for any other production use (Licensing). A release is therefore supplied in the course of a commercial activity, a free download included.
- Not free and open-source software. Art 3(48) asks for “a free and open-source licence which provides for all rights to make it freely accessible, usable, modifiable and redistributable”. The Business Source License 1.1 withholds production use outside its Additional Use Grant, so FerroFED is not free and open-source software within the CRA, and the open-source software steward of Art 3(14) and Art 24 does not apply to it.
- The same test under the EHDS. EHDS Art 2(1)(d) takes “placing on the market”, “manufacturer” and “economic operator” from Regulation (EU) 2019/1020, whose Art 3(1) has the same words: “in the course of a commercial activity, whether in return for payment or free of charge”.
What the manufacturer does
- One product per release. Every tagged release, meaning the source tag,
the binary tarballs, the gateway image and the console image of one
version, is one product with digital elements. Cadasto B.V. places it on
the market as its manufacturer (Art 3(13)). The
mainbranch is development code and is not supplied for use. - Reporting applies now. Art 14 has the manufacturer notify an actively
exploited vulnerability and a severe incident, with an early warning
within 24 hours. Art 71(2) applies Art 14 “from 11 September 2026”, and
Art 69(3) applies it to “all products with digital elements … placed on
the market before 11 December 2027”, the v0.0.x releases among them. The
written reporting procedure is in
SECURITY.mdand on the Complaints and incidents page (#762). - The rest from 11 December 2027. Art 71(2) applies the rest of the CRA from 11 December 2027. Each release placed on the market from that date meets the essential requirements of Annex I and has the support period below. Art 69(2) holds a product placed before that date to those requirements only once it is “subject to a substantial modification”.
- One file, one declaration, one CE marking. EHDS Art 104 inserts CRA Art 32(5a): the manufacturer of an EHR system shows conformity with CRA Annex I “using the relevant conformity assessment procedure provided for in Chapter III” of the EHDS Regulation. It also replaces CRA Art 31(3), so “a single set of technical documentation shall be drawn up” for both acts. CRA Art 28(3) and EHDS Art 39(2) each require a single EU declaration of conformity. A release placed from 11 December 2027 therefore carries one technical documentation set, one declaration and one CE marking for both acts, all planned under #525.
- The economic operators supplied. EHDS Art 35 has an economic operator identify, for 10 years, “any economic operator to which they have supplied an EHR system”. CRA Art 23(1)(b) asks the same “where available”; EHDS Art 35 has no such qualifier. The manufacturer’s answer is a register, kept for 10 years, of every economic operator it supplies under a contract: commercial licensees, resellers, integrators and hosts. The procedure is in place (The register of economic operators): the register’s format is public and checked in CI, and its rows are kept by the manufacturer outside the public repository and are not published. The public release is not gated, and it is supplied for non-production use and for non-commercial production use under the licence. If counsel holds that an anonymous production user is an economic operator, for example one that puts FerroFED into service under EHDS Art 26(2), downloads move behind a registration.
A FerroFED that someone hosts as a service, or that a health institution builds and runs for itself, is outside the CRA for that service: EHDS recital 112 says that “EHR systems offered through the SaaS licensing and delivery model do not fall within the scope of” the CRA, nor do “EHR systems that are developed and used in-house”. EHDS Art 26(2) still counts both as put into service (A deployment a health institution runs for itself). The release that was downloaded to run it stays a CRA product.
Two delegated acts are adopted under the CRA, both vendored beside it. Commission Delegated Regulation (EU) 2025/1535 excludes from the CRA the products within the scope of Regulation (EU) No 168/2013, on two- or three-wheel vehicles and quadricycles (its Art 1); FerroFED is not one of them, so the act changes nothing here. Commission Delegated Regulation (EU) 2026/881, adopted under Art 14(9), sets when the CSIRT that first receives a notification may delay passing it on (Art 16(2)). It binds that CSIRT and puts no duty on the manufacturer; the Complaints and incidents page says what it means for a notification.
The support period
Art 13(8) has the manufacturer handle a product’s vulnerabilities for its
support period, and “the support period shall be at least five years”. Art
13(19) has the end date of that period, “including at least the month and
the year”, specified “at the time of purchase”, and Annex II, point 7, puts
“the end-date of the support period” in the information to the user.
Cadasto B.V. meets this as a commitment for the whole product: for every
release placed on the market from 11 December 2027 it keeps publishing
security fixes, in new releases, for at least five years, with the end date
published, and ships them only in the newest release, as Art 13(10) allows.
Only the newest release is supported: when a new release is published,
every earlier release stops receiving fixes, and upgrading is free. There
are no maintenance branches, no long-term-support line and no backports.
SECURITY.md
lists every release with the date until which fixes are published, five
years from its release date
(#763). The gateway
knows the end date of its own release from its build: the startup banner and
OPTIONS {base}/ name it, and once it has passed the banner, the log and
config check say so
(#782).
The basis of the five years
Art 13(8), second subparagraph, has the manufacturer “determine the support period so that it reflects the length of time during which the product is expected to be in use, taking into account, in particular, reasonable user expectations, the nature of the product, including its intended purpose, as well as relevant Union law determining the lifetime of products with digital elements”. Its fifth subparagraph puts “the information that was taken into account” in the technical documentation, which Annex VII, point 4, repeats. Cadasto B.V. decided on 2026-10-16 to keep five years from each release date (#1088). It took these matters into account:
- The length of time in use. A gateway sits between the hospital platforms it federates: the CDRs behind it and the clients in front of it. The manufacturer expects one release to stay in use until those platforms are next upgraded, when the operator moves the gateway to a newer release as part of the same change, and expects that interval to fall within five years.
- Reasonable user expectations. The users are the health institutions
and the operators that run the gateway for them
(Intended purpose). They can expect every
vulnerability found to be fixed in a new release they can take free of
charge: a security fix ships as a patch release of the newest release,
with the fix alone, or with the reason a functional change ships with it
stated in the release notes
(
SECURITY.md). - The nature of the product and its intended purpose. FerroFED is software with no hardware part, and it holds no clinical data of its own, so an operator replaces a release by installing a newer one (Upgrading).
- Union law determining the lifetime. Art 13(8), third subparagraph, sets the floor: “the support period shall be at least five years”. The acts adopted under the CRA that the manufacturer holds, Regulations (EU) 2025/1535, 2025/2392 and 2026/881, set no support period. EHDS Art 39(4) keeps a digital declaration of conformity online “for the expected lifetime of the EHR system and, in any event, for at least 10 years”; that rule sets how long the declaration stays available, and sets no support period.
- The release cadence. Releases come days or weeks apart: seven between
2026-10-01 and 2026-10-15, in the table of
SECURITY.md. Only the newest release is supported, and each release starts its own five years, so fixes keep coming for at least five years after the newest release.
Art 13(8) also lets the manufacturer take into account “the support periods of products with digital elements offering a similar functionality placed on the market by other manufacturers, the availability of the operating environment, the support periods of integrated components that provide core functions and are sourced from third parties as well as relevant guidance provided by the dedicated administrative cooperation group (ADCO)”. Of these:
- Similar products. The manufacturer compared no other gateway’s support period, and the period does not rest on one.
- The operating environment and integrated components. A release runs as its container image or binary tarball and reaches each node over openEHR ITS-REST. A vulnerability in a component the release carries, which its SBOMs list, is fixed in a new release like one in the gateway’s own code, so no supplier’s end of support ends the commitment.
- ADCO and Commission guidance. The manufacturer holds no ADCO or Commission guidance on support periods (Art 52(16)) for this category of product, and reads the period against such guidance once it is published.
Already in place
None of these is a conformity claim; each answers a CRA requirement in part:
- Each release attaches a CycloneDX SBOM of the source dependency graph and an SPDX SBOM of the shipped binary, with signed provenance (Annex I, Part II, points 1 and 7).
- Vulnerabilities are reported privately, with an acknowledgement within about 5 business days and an initial assessment within about 10, and a fixed release follows within 7, 30 or 90 days of the confirmed assessment for a Critical, High or Medium vulnerability, and with the next release for a Low one (Annex I, Part II, points 5 and 6; Art 13(17)).
- The running system names the manufacturer, its postal address and its single point of contact (Art 13(16)).
- A published release cannot be changed or deleted, so every update stays available (Art 13(9)).
- The cybersecurity risk assessment reads the threat model and the clinical safety risk file against Annex I (Art 13(2), (3)), and CRA information to the user gives the points of Annex II (Art 13(18)).
What is built and what is planned
Neither harmonised software component is complete yet. Three built features relate to the essential requirements, and none has been assessed against them:
- Every caller is authenticated at the gateway (client authentication). Annex II, point 3.1, asks for “reliable mechanisms for the identification and authentication of health professionals”. For patient data the gateway requires a natural person behind the token, or a client acting for the professional the token names, and the assurance level the deployment sets per issuer (Professionals and assurance).
- Every access to patient data the gateway intermediates is recorded with
the verified caller, the patient, each endpoint asked and the categories
of the data, in the record Annex II, point 3.2, sets for the European
logging component (the access log).
The library that builds it is
ehds-logging, published from the FerroHEALTH audit library (FerroAUDIT). The records are reviewed at the Audit Record Repository with ITI-81, the connection of external software point 3.3 admits, and each states how long it is kept by its origins and categories, never under three years (point 3.4, Art 9(2); Reading the log). A request addressed byehr_idnames the patient the identity binding holds under it, so a search by the person’s identifier finds it (Art 9(1)); a query over many patients’ data names the patient behind eachehr_idits delivered rows project, up to a bound, and otherwise records its scope, which a person’s search finds through the members that hold their EHR (A query over many patients’ data). Each record carries a security label per category and per origin, which the repository applies as access rights by origin and category (point 3.4; Labelling the access records). - The European interoperability component is the
eehrxfcrate, published from FerroHEALTH/FerroBRIDGE and pinned exactly (its pages): the EHDS dataset model, read from the Xt-EHR logical models, and the mapping of an openEHR composition to a FHIR R4Bundlethrough the mapping files you supply. It also reads a received document, checks it against its profiles and maps it into one openEHR composition that keeps the original document (Receiving a document). The gateway holds the stored queries that select each patient summary section’s compositions (the section queries), and serves the patient summary built from them on a FHIR R4 face of its own, the interface of Annex II, point 2.1 (The patient summary over FHIR). The same face lists the summaries that exist for a patient, the members’ own ones included (The document list). It also searches and reads a patient’s electronic prescriptions and receives a dispensation into the member that owns its prescription, passed through in the HL7 Europe MPD profiles from the members that serve them on their own FHIR endpoint, or mapped from a member’s openEHR medication data in the map mode (Prescriptions and dispensations over FHIR). It lists, reads and receives laboratory reports in the HL7 Europe Laboratory Report profiles (Laboratory reports over FHIR), lists, reads and receives discharge reports in the Hospital Discharge Report profiles (Discharge reports over FHIR), and lists and reads imaging reports in the Imaging Report profiles (Imaging reports over FHIR).
Seven of the manufacturer’s obligations have a first answer:
- The running system names its manufacturer, with the postal address and the
single point of contact (Art 30(1)(g)):
GET {base}/,OPTIONS {base}/, the startup banner,ferrofed --version, the operator console’s About page, which the footer of every console page links to, and the image labels. - The complaint channels, the two registers, corrective action and the serious-incident report (Art 30(1)(i) to (o), Art 44(7)) are written procedures (Complaints and incidents).
- The economic operators supplied under a contract are identified from a register kept for 10 years outside the public repository (Art 35, CRA Art 23), with its procedure written and its public format checked in CI (The register of economic operators).
- Every public text is reviewed against Art 28 at each release, and the completeness default, the budgets and the consent pre-filter are assessed against Annex II, point 2.5 (Claims review).
- The information sheet gives every item of Art 38(2), held to the pin matrix by a guard, and the instructions for use cover installation, the configuration of a deployment in the EU, maintenance and the limitations (Art 30(1)(d), Art 32(2)(d), Annex III, point 1(j)).
- The clinical safety risk file names each hazard of the two harmonised components with its control and the test that holds it, for Annex II, point 1.1.
- The technical documentation indexes every element of Annex III, a release that changes a harmonised component is not cut without a review row (Art 30(2)), and each release’s documentation is kept for ten years (Art 30(3)).
The planned work is filed under #519. What remains before the Regulation applies on 26 March 2027:
- #799: the European logging software component is built (#521), and is re-checked against the Art 36(1) common specifications once they are adopted. None is adopted, so the issue waits upstream with no milestone.
- FerroHEALTH/FerroBRIDGE#1327: the European interoperability software component for the patient summary is built from federated openEHR data (#522), and is re-checked against the Art 15(1) implementing act once it is adopted. None is adopted.
- #1794: the rating of what a received discharge or imaging report can lose. The component moved to FerroBRIDGE (#813), where prescriptions from openEHR medication data, laboratory reports, discharge reports and imaging were built (issues FerroHEALTH/FerroBRIDGE#1321, FerroHEALTH/FerroBRIDGE#1322, FerroHEALTH/FerroBRIDGE#1324, FerroHEALTH/FerroBRIDGE#1325 and FerroHEALTH/FerroBRIDGE#1326, now closed).
- #525: the rest of the conformity work: the points the Annex II checklist leaves open, the EU declaration of conformity, which each release writes once the manufacturer issues it, the CE marking and the registration.
No EU declaration of conformity has been drawn up and no FerroFED release
carries the CE marking. The release process says when, in a release, the
manufacturer affixes the CE marking (Art 41) and makes or updates the EU
database entry (Art 49), and that the entry waits on the list of data the
Art 49(4) delegated act will set
(docs/release.md,
Before the tag, step 8).
The crosswalk of each category, which ties the eHealth Network dataset, the Xt-EHR logical model and the HL7 Europe profiles together, and the section queries that select the openEHR content, are on The category crosswalks.
Not legal advice
This page is the manufacturer’s reading of the EHDS Regulation and the CRA, written to plan the engineering work. It is not legal advice, and it does not decide any deployment’s obligations. The manufacturer’s own counsel is asked to confirm, now, that a release is supplied in the course of a commercial activity, that Cadasto B.V. is its manufacturer, and that CRA Art 14 applies to the v0.0.x releases; the rest of the CRA position is put to counsel before the first release placed on the market from 11 December 2027. A deployment’s counsel should confirm at least:
- The intended-purpose statement and the classification as an EHR system under Art 2(2)(k), including the answer to recital 38’s “general-purpose middleware” example.
- That the intended purpose covers all six priority categories, given that FerroFED passes through whatever the member CDRs hold.
- Whether the deployment is an EHR system “manufactured and used within health institutions” (Art 26(2)), and who then carries the manufacturer’s obligations of Art 30.
- Whether the deployment changes FerroFED in a way that makes its operator a manufacturer under Art 34.
- Whether hosting FerroFED for others is offering it as a service under Art 26(2). The manufacturer reads each tagged release as placed on the market by Cadasto B.V. (The Cyber Resilience Act; Art 2(1)(d) takes the definition from Regulation (EU) 2019/1020). What remains for a deployment is whether it is an economic operator that Cadasto B.V. supplied, which EHDS Art 35 has the manufacturer identify: 2019/1020 Art 3(13) counts whoever puts a product into service under the applicable legislation, and EHDS Art 26(2) counts a hosted or in-house system as put into service.
- The dates for the Chapter III articles Art 105 does not name, read together with Art 26(1).
- The national rules that apply beside the Regulation: national requirements for EHR systems (Art 42) and the Member State’s arrangements for connecting healthcare providers to its national contact point (Art 23(5)).